MDCX Series Pressure Sensors
-
Features
- 0 to 1” H2O to 0 to 30” H2O Pressure Ranges
- Matched pressure port volumes
- Temperature Compensated
- Calibrated Zero and Span
-
Applications
- Medical Instrumentation
- Environmental Controls
- HVAC
equivalent Circuit

Pressure Sensor Maximum Ratings
Environmental Specification
Help Me Choose
Choose A Part Number Below To View Performance Characteristics
2 INCH-GDIP-MV-MINI Performance Characteristics
ClosePARAMETER | MIN | NOM | MAX | UNITS | Notes |
Operating Range, differential pressure | - | 2.0 | - | inH2O | - |
Output Span, @ 2 inH2O | 9.0 | 10.0 | 11.0 | mV | 5 |
Operating Range, differential pressure | - | 4.0 | - | mV | - |
Output Span, @ 4 inH2O | 18.0 | 20.0 | 22.0 | mV | 5 |
Offset Voltage @ zero differential pressure | - | - | ±500 | uV | - |
Offset Temperature Shift (0°C-50°C) | - | - | ±250 | uV | 2 |
Offset Warm-up Shift | - | - | ±100 | uV | 3 |
Offset Position Sensitivity (1g) | - | - | ±50 | uV | - |
Offset Long Term Drift (one year) | - | - | ±200 | uV | - |
Linearity, hysteresis error | - | 0.05 | 0.25 | %fs | 4 |
Span Shift (0°C-50°C), @ 2 inH2O | - | - | ±200 | uV | 2 |
Standard Pressure Range
Operating Pressure
0 - 2 inH2O
Specification notes
NOTE 1: ALL PARAMETERS ARE MEASURED AT 12.0 VOLT EXCITATION, FOR THE NOMINAL FULL SCALE PRESSURE AND ROOM TEMPERATURE UNLESS OTHERWISE SPECIFIED. PRESSURE MEASUREMENTS ARE WITH POSITIVE PRESSURE APPLIED TO THE B-PORT CONFIGURATION.
NOTE 2: SHIFT IS RELATIVE TO 25°C.
NOTE 3: SHIFT IS WITHIN THE FIRST HOUR OF EXCITATION APPLIED TO THE DEVICE.
NOTE 4: MEASURED AT ONE-HALF FULL SCALE RATED PRESSURE USING BEST STRAIGHT LINE CURVE FIT.
NOTE 5: THE VOLTAGE ADDED TO THE OFFSET VOLTAGE AT FULL SCALE PRESSURE.