AXCX Series Prime Grade Medium Pressure

The Millivolt Output pressure sensors is based upon a proprietary packaging technology to reduce output offset or common mode errors. This model provides a calibrated millivolt output with excellent output offset characteristics. In addition the sensor utilizes a silicon, micromachined, stress concentration enhanced structure to provide a very linear output to measured pressure.

These calibrated and temperature compensated sensors give an accurate and stable output over a wide temperature range. This series is intended for use with non-corrosive, non-ionic working fluids such as air, dry gases and the like. The PRIME GRADE is the highest accuracy version of the millivolt output pressure sensors. The output of the device is ratiometric to the supply voltage and operation from any D.C. supply voltage up to +16 V is acceptable.

  • Features

    • 0.3 to 150 PSI Pressure Ranges
    • Highest Accuracy Version
    • Temperature Compensated
    • Calibrated Zero and Span
  • Applications

    • Medical Instrumentation
    • Environmental Controls
    • HVAC

equivalent Circuit

Pressure Sensor Maximum Ratings

Supply Supply Voltage VS

16 Vdc

Common-mode pressure

50 psig

Lead Temperature (soldering 2-4 sec.)

270°C

Environmental Specification

Compensated

0 to 70°C

Operating

-25 to 85°C

Storage

-40 to 125°C

Humidity Limits

0 to 95% RH (non condensing)

Available Package

Click On Package Image Below To View Package Drawing

Help Me Choose

Choose A Part Number Below To View Performance Characteristics

Pressure Range Differential Absolute Gauge
0.3 PSI 0.3 PSI-D-PRIME-MV -- --
1 PSI 1 PSI-D-PRIME-MV -- --
5 PSI 5 PSI-D-PRIME-MV -- --
15 PSI 15 PSI-D-PRIME-MV 15 PSI-A-PRIME-MV --
30 PSI 30 PSI-D-PRIME-MV 30 PSI-A-PRIME-MV --
100 PSI 100 PSI-D-PRIME-MV 100 PSI-A-PRIME-MV --
150 PSI 150 PSI-D-PRIME-MV -- --
 

30 PSI-A-PRIME-MV Performance Characteristics

Close
PARAMETER MIN NOM MAX UNITS Notes
Output Span 99.0 100 101 mV 4
Offset Voltage @ zero absolute pressure - - ±0.3 mV -
Offset Temperature Shift (0°C-70°C) - - ±250 uV 2
Linearity, hysteresis error - 0.15 0.30 %FSS 3
Span Temperature Shift (0°C-70°C) - - ±1.0 %FSS 2

Standard Pressure Range

Operating Pressure

0 - 30 PSIA

Nominal Span

90 mV

Proof Pressure

90 PSIA

Burst Pressure

150 PSIA

Specification notes

NOTE 1: ALL PARAMETERS ARE MEASURED AT 12.0 VOLT EXCITATION, FOR THE NOMINAL FULL SCALE PRESSURE AND ROOM TEMPERATURE UNLESS OTHERWISE SPECIFIED. PRESSURE MEASUREMENTS ARE WITH POSITIVE PRESSURE APPLIED TO PORT B.


NOTE 2: SHIFT IS RELATIVE TO 25°C.


NOTE 3: MEASURED AT ONE-HALF FULL SCALE RATED PRESSURE USING BEST STRAIGHT LINE CURVE FIT.


NOTE 4: THE SPAN IS THE ALGERBRAIC DIFFERENCE BETWEEN FULL SCALE OUTPUT VOLTAGE AND THE OFFSET VOLTAGE.

Standard Package