ELVR Series
-
Features
- 2.5 to 75 mbar Pressure Ranges
- Digital I2C or SPI Interfaces with Analog Output
- 12 Bit Digital Resolution
- 3V or 5V Supply Voltage Options
- All Sensors' Superior, Dual-Die CoBeam2 TM Technology
- Custom Pressure Range and Calibration Outputs Available Upon Request
-
Applications
- Medical
- Industrial
equivalent Circuit

Pressure Sensor Maximum Ratings
Environmental Specification
ELVR-L20G Performance Characteristics
ClosePARAMETER | MIN | NOM | MAX | UNITS | Notes |
Full Scale Span (FSS) | 3.9 | 4.0 | 4.1 | V | 5 |
Full Scale Output | - | - | - | - | - |
- Min Pressure | - | 0.25 | - | V | - |
- Max Pressure | - | 4.25 | - | V | - |
Span Temperature Shift | - | - | ±1 | %FSS | 2 |
Offset Voltage @ zero differential pressure | 0.15 | 0.25 | 0.35 | V | - |
Offset Temperature Shift | - | - | ±0.5 | %FSS | 2 |
Offset Warm-up Shift | - | ±5 | - | mV | 3 |
Offset Long Term Drift (one year) | - | ±5 | - | mV | - |
Linearity, Hysteresis Error | - | ±0.05 | ±0.25 | %FSS | 4 |
Response Delay | - | 0.5 | - | ms | 1 |
D/A Resolution | - | - | 11 | bit | - |
Current Consumption | - | 5 | - | mA | 6 |
Standard Pressure Range
Operating Range
50 mbar
Proof Pressure
500 mbar
Burst Pressure
1250 mbar
Specification notes
NOTE 1: For any pressure applied the optput response time to reach 90% of pressure applied.
NOTE 2: Shift is relative to 25°C between standard or industrial compensated temperature range endpoints.
NOTE 3: Shift is within the first hour of excitation applied to the device.
NOTE 4: Measured at one-half full scale rated pressure using best straight line curve fit.
NOTE 5: The span is the algebraic difference between full scale output voltage and the offset voltage.
NOTE 6: Parameter is characterized and not 100% tested.