DLV Series

The DLV Series Mini Digital Output Pressure Sensor is based on the already popular DLVR series pressure sensors. This series offers excellent performance over middle pressure ranges of 5 to 60 PSI, compared to the DLVR low pressure range products.

 

The supply voltage options ease integration of the sensors into a wide range of process control and measurement systems, allowing direct connection to either I2C or SPI serial communications channels. For battery-powered systems, the sensors can enter very low-power modes between readings to minimize load on the power supply

 

These calibrated and compensated sensors provide accurate, stable output over a wide temperature range. This series is intended for use with non-corrosive, non-ionic working fluids such as air, dry gasses and the like. A SIL-Gel die coating is added for enhanced media protection. 

  • Features

    • 5 to 60 PSI and Barometric Pressure Ranges
    • 3.3V or 5.0V Supply Voltage
    • I2C or SPI Interface
    • High Resolution 14 Bit Output
    • Better than 0.5% Accuracy Over Temperature Typical
  • Applications

    • Medical Breathing
    • Enviromental Controls
    • HVAC
    • Industrial Controls
    • Portable/Hand-Held Equipment

equivalent Circuit

Pressure Sensor Maximum Ratings

Supply Voltage, Vs

6 Vdc

Common Mode Pressure

10 psig

Lead Temperature (soldering 2-4 sec.)

270°C

Environmental Specification

Compensated Operating

-

Commercial

0 to 70°C

Industrial

-20 to 85°C

Operating

-25 to 85°C

Storage

-40 to 125°C

Humidity Limits (non condensing)

0 to 95% RH

Available Package

Click On Package Image Below To View Package Drawing

Help Me Choose

Choose A Part Number Below To View Performance Characteristics

Pressure Range Differential Absolute Gauge
5 psi DLV-005D -- DLV-005G
15 psi DLV-015D DLV-015A DLV-015G
30 psi DLV-030D DLV-030A DLV-030G
60 psi DLV-060D DLV-060A DLV-060G
600 to 1100 mBarA -- DLV-611M --
 

DLV-005D Performance Characteristics

Close
PARAMETER MIN NOM MAX UNITS Notes
Output Span - ±6,553 - Dec count 1
Offset Output - 8,192 - Dec count -
Total Error Band - ±0.5 ±1.0 %FSS 2
Span Temperature Shift - ±0.1 - %FSS 3
Offset Temperature Shift - ±0.1 - %FSS 3
Offset Warm-up Shift - ±0.1 - %FSS 4
Offset Long Term Drift (One Year) - ±0.1 - %FSS -
Linearity, Hysteresis Error - ±0.1 ±0.25 %FSS 6
Response Delay - - - - 5,10
Sleep-Wake Pressure - 0.40 0.50 ms 5
Sleep-Wake All - 1.10 1.40 ms 5
Power-On to First Reading Attempt 6.0 + 1 update period - - ms -
Update Rate - - - - -
Fast - 0.40 1.0 ms 5
Noise Reduced - 1.30 3.1 ms 5
Low Power - 6.5 9.5 ms 5
Start-up Time - - 10.0 ms 5,7
Digital Resolution - - - - -
Output Resolution - 14 - bit 5
No Missing Codes 12 13 - bit 5
Temperature Output - - - - -
Resolution - 11 - bit 8
Overall Accuracy - 2 - °C 8
Current Requirement (3.3V Option) - - - - -
Fast - 2.5 3.2 mA 5
Noise Reduced - 2.5 3.1 mA 5
Low Power - 0.6 0.7 mA 5
Sleep (Idle) - 0.5 5.0 uA 5
Current Requirement (5.0V Option) - - - - -
Fast - 3.5 4.5 mA 5
Noise Reduced - 3.6 4.4 mA 5
Low Power - 0.8 1.0 mA 5
Sleep (Idle) - 0.5 5.0 uA 5

Standard Pressure Range

Operating Range

±5 PSI

Proof Pressure

20 PSI

Burst Pressure

40 PSI

Nominal Span

±6,553 counts

Specification notes

NOTE 1: The span is the algebraic difference between full scale decimal counts and the offset decimal counts.

 Pressure Output Transfer Function:

                                 

 Where,

               Poutdig                            Is the sensor 14 bit digital output

               OSdig                              Is the specified digital offset (gage and absolute = 1,638)

               FSS (psi)                        Is the sensor Full Scale Span in psi

 

NOTE 2: Total error band comprises of offset and span temperature and calibration errors, linearity and pressure hysteresis errors, offset warm-up shift, offset position sensitivity and long term offset drift errors.

NOTE 3: Shift is relative to 25°C.

NOTE 4: Shift is within the first hour of excitation applied to the device.

NOTE 5: Parameter is characterized and not 100% tested.

NOTE 6: Measured at one-half full scale rated pressure using best straight line curve fit.

NOTE 7: Power-on time is time from power being applied to first available part communications.

NOTE 8: Temperature output conversion function:

                          Temperature Output Transfer Function: